Darkroom Recommendations For X-ray Film Processing?

Sep 02, 2025 Leave a message

Cleaning the processing tank: Whenever the processing solution is changed, the tank must be cleaned, preferably with hot water and soap. If this is insufficient, polyester tanks can be cleaned with a bleach solution (100-200 ml per liter of water), hydrochloric acid (10 ml per liter of water), or acetic acid (50 ml per liter of water). Stainless steel tanks can be cleaned with a nitric acid solution (10 ml per liter of water) or acetic acid (50 ml per liter of water). However, hydrochloric acid should never be used to clean stainless steel tanks.

There are also industrial cleaners available specifically for darkroom cleaning, such as Devclean and the environmentally friendly Fixclean.

Contaminated fingers: After contact with the developer, rinsing with water can prevent brown stains on fingers. If fingers are truly contaminated, they should be immersed in the following solutions:
a. 1 liter of water
b. 2 grams of potassium permanganate
c. 10 ml of concentrated sulfuric acid
Hands should then be rinsed in an acidic fixer solution and finally washed with soap and water.

Chalked Water:
Using hard, chalky water to mix solutions can cause troublesome handling problems. Calcium salts in the presence of carbonates and sulfites can form a water-insoluble white deposit on the film. To prevent this, soften the diluent using a special filter or boil and cool it before preparing the chemical solution.
To remove chalky deposits from film, soak it in a solution of 7 ml of glacial acetic acid per liter of water.

Send Inquiry

whatsapp

Phone

E-mail

Inquiry